A digestion turns a solid or a slurry into a solution an instrument can read. The route the buyer chooses decides how much acid the buyer buys each year, how long an analyst stands at the bench, and whether mercury or arsenic survives the process at all.
Three routes open a sample: an open vessel on a hotplate, a sealed vessel in a microwave, and a large tube in a heated block. They are not three grades of the same thing. Each one fails on a different matrix, and the failure is usually discovered months after installation.
The instrument sets the detection limit the buyer can claim. The digestion decides whether the buyer ever reach it.
Ask two questions before anything is quoted. Does the analyte survive heat in open air, and how large is the reagent blank against the level the buyer has to report.
Mercury, arsenic and selenium are the classic losses. In an open vessel they leave with the vapour, and how much leaves depends on how long the analyst refluxed and how often the acid was topped up. A sealed vessel keeps them in solution.
The second question is less obvious and more expensive. When the blank sits close to the reporting limit, a better detector reads the blank more precisely and changes nothing. The remedy is the acid and the vessels.
An open tube on a hotplate is the least expensive thing to buy and the most expensive thing to run. The acid boils off and has to be replaced, and the replacement is reagent cost that never stops. Somebody also has to stand there and watch it.
HYSI does not catalogue an open hotplate. When a customer describes that workflow, it is usually because a method was written around it years ago and the method, rather than the equipment, is the thing worth revisiting first.
Two running costs are easy to miss at the specification stage. Acid consumption scales with the number of positions, and the fume extraction has to cover every one of them.
A sealed vessel lets temperature and pressure do the work instead of the analyst. The microwave digestion system in the catalogue holds 12 or 16 PTFE-TFM vessels, reaches 260 °C and 15 MPa, and delivers 1800 W. Programs are stored on a touchscreen, so the same method runs the same way on a Monday and on a Friday.
Two habits keep the route repeatable. Run a reagent blank in one position of every batch, and keep the vessel set matched to the rotor, because a mixed set heats unevenly.
Vessel count is a throughput decision, not a purchasing decision. Buying exactly the batch size the buyer runs today means the first damaged vessel stops the laboratory for a week.
A heated block is the route for large tubes, high temperature and no pressure. The Kjeldahl digestion unit in the catalogue takes 20 tubes of 250 mL, holds ambient to 450 °C on an aluminium block, and carries an integrated exhaust manifold with digital PID programs.
This is the route for nitrogen determination, and for methods that need a large sample mass or a large acid volume. There is no pressure vessel to maintain, which is why it stays in laboratories that also own a microwave.
A block does not decompose refractory silicates. That remains a sealed-vessel job.
Trace work is limited by the blank, and the blank is limited by the acid.
An acid purification and distillation system takes analytical grade acid and redistils it, with quartz and PTFE on every wetted surface. The catalogue figure is metal impurities below 10 ppt in HNO3, at 1000 mL per batch.
Buy this before the second instrument, not after.
| Route | How the attack is driven | Batch per run | Where it fits | Main risk |
|---|---|---|---|---|
| Open vessel on a hotplate | Reflux at atmospheric pressure, acid topped up by hand | Set by bench space | Legacy methods, low sample numbers | Volatile analytes leave; acid consumption |
| Sealed vessel, microwave | Temperature and pressure, 260 °C and 15 MPa | 12 or 16 vessels | Refractory matrices, volatile analytes, trace work | Vessel condition, matched sets |
| Heated block, large tube | Block to 450 °C at atmospheric pressure, manifold extraction | 20 × 250 mL tubes | Nitrogen, large sample mass, large acid volume | Silicates stay undecomposed |
Send the matrix, the analyte list, the level the buyer has to report, the number of samples per day, and the method the buyer has to follow if one is already fixed. That is enough for HYSI to name a route and a model.
The manufacturer answers with the model, the part number and the delivery term, with the vessel sets and the acid as separate items so the cost of each is visible. If an open hotplate is genuinely right for the method, the page says so.
For trace work with volatile analytes it is the better route, because the vessel holds what an open reflux loses. If the buyer's method fixes an open reflux in writing, the method has to be revised before the equipment changes.
More than one batch. A vessel set is a consumable with a service life, and a batch that runs exactly at the vessel count stops the laboratory the first time one is damaged.
Because the blank arrives with the reagent, not with the sample. If it sits near the level the buyer has to report, a better detector reads the blank more precisely and changes nothing.
Only when the matrix contains silicate. Ask which liner the vessel carries before the method is written, because HF changes the vessel, the handling and the fume extraction.
It is the least expensive way to move a reporting limit that is set by the blank. Buy it before the second instrument.