Temperature uniformity
Conflicting温度均一性 国标用词:温度均匀度
One machine, two notations: ±3 °C against a centre point, or 3 °C across the whole chamber. The two figures look like a factor of two and can describe the same unit.
同一台机器,国际口径报「±3 °C」(相对中心点)、中国标准报「3 °C」(全空间跨度)。两个数字碰在一起,看起来差一倍,其实可能是同一件事。
What it measures定义什么
How far temperature differs between positions inside the chamber at the same instant. It answers whether the top-left corner sits at the same temperature as the bottom-right, and says nothing about time.
同一时刻,箱内不同位置的温度差异。它回答的是「左上角和右下角是不是同一个温度」,与时间无关。
How it is measured怎么测
Multi-point mapping, commonly nine points, read at one instant once the unit is in steady state. The international convention publishes the maximum deviation from a centre reference as ±X °C; the Chinese standard publishes upper deviation minus lower deviation as one span, alongside the pair.
多点布测(品牌白皮书常用 9 点),在稳定状态下同一时刻读数。国际惯例报「相对中心参考点的最大偏差」±X °C;中国标准报「上偏差值与下偏差值之差」,并同时给出上偏差/下偏差一对数。
Where the numbers part口径差异
One word, two arithmetic conventions, figures differing by nearly a factor of two. ±3 °C against a centre point implies a span of up to 6 °C; the Chinese standard (NIFDC 3.38) publishes upper minus lower, which is that span. Set side by side, ±3 reads as better than 3 while describing the same unit; equated, 3 against ±3 doubles the span. A second and less-mentioned factor: inside one series a larger chamber reports a larger number. Binder UF V 500 (501 L) publishes 2.5 ± K, and UF V 350 (402 L) publishes 3.5 ± K, both to DIN 12880, empty, at 22 °C ± 3 °C. Uniformity therefore tracks chamber size, so comparing across volumes misleads as much as comparing across makes.
同一个词,两种算法,数值差近两倍。国际口径报「相对中心点最大偏差 ±3 °C」,意味着全空间跨度最大 6 °C。中国标准口径(NIFDC 3.38)报「上偏差 − 下偏差」,其值就是全空间跨度 3 °C。把两个数字并排放,±3 会被误读成比 3 好,实际是同一件事的两种写法;反过来把 3 与国际的 ±3 等同,就等于把跨度高估一倍。 另有一层更少被提到的差异:**同一品牌、同一系列,容积不同,数字可以差两三倍。** Binder UF V 500(501 L)报 2.5 ± K / 波动 1.5 ± K,而 UF V 350(402 L)报 3.5 ± K / 波动 4.0 ± K,两者测法完全相同(DIN 12880、空箱、22 °C ± 3 °C)。所以均匀度是随箱体尺寸走的,跨容积比数字没有意义——这一点比跨品牌比更容易被忽略,因为同一个系列看起来「应该一样」。
Ask before comparing看数字前先问
看到均匀度数字,先问三件事:是 ±单值还是一对上下偏差?基准是中心点还是设定值?设定点与环境温度是多少?三项未齐,数字不可比。
Sources 来源- PHCbi · Ultra-Low Temperature Freezers: By the Numbers (white paper) · 原文Tier 2 · manufacturer technical document
- Binder · Model UF V 500 technical data(官方产品页) · 原文Tier 2 · manufacturer technical document
- Binder · Model UF V 350 technical data(官方产品页) · 原文Tier 2 · manufacturer technical document
- PHCbi · ULT freezer brochure(厂家手册 PDF,经第三方镜像取得) · 原文Tier 2 · manufacturer technical document
- NIFDC (National Institutes for Food and Drug Control) · 医用冷链设备术语(公开征求意见稿)3.38 · 原文Tier 4 · standard or regulator
- Thermo Fisher Scientific · Thermo Scientific TSX 系列变频超低温冰箱 白皮书(中文版,品牌自有域名) · 原文Tier 2 · manufacturer technical document
- Thermo Fisher Scientific · 超低温冰箱的特点(品牌 ULT 特性页,中文版) · 原文Tier 2 · manufacturer technical document
- Thermo Fisher Scientific · TSE240V Technical Specification Sheet(品牌官方数据表) · 原文Tier 3 · industry guidance
- Haier Biomedical · -40°C Biomedical Freezer DW-40L298J / 418D / 528D 彩页 · 原文Tier 2 · manufacturer technical document
Temperature fluctuation
Differs温度稳定性 国标用词:温度波动度
English separates fluctuation from stability; Chinese brand pages collapse both into one word. Given ±0.5 °C, ask whether it is the swing inside a control cycle or drift across a year.
英文两个词(fluctuation / stability),中文只有一个「稳定性」。看到 ±0.5 ℃,先问是控制周期内的起伏,还是长期漂移。
What it measures定义什么
The rise and fall of temperature over time at a single measuring point. It answers whether that point holds steady, and says nothing about spatial distribution.
同一个测点,温度随时间的上下起伏。它回答的是「这一个点稳不稳」,与空间分布无关。
How it is measured怎么测
Continuous sampling across one or more control cycles in steady state, taking the difference between the highest and lowest instantaneous reading. The Chinese standard computes this at every point and takes the largest difference; a datasheet usually gives only the point where the control probe sits.
在稳定状态下连续采样一个或多个温度控制周期,取单点瞬时最高与最低之差。中国标准要求对箱内每个测点都算,取其中最大的差值;品牌数据表通常只给控制探头所在的那一个点。
Where the numbers part口径差异
The scope differs, not merely the number. The Chinese standard takes the worst point in the chamber; a datasheet often reports the control probe alone, so the two can differ by more than a factor of two while looking like the same indicator. A further trap: some documents treat stability as a synonym for fluctuation, others mean long-term annual drift, and the two are not interchangeable. One counter-intuitive case: Binder UF V 350 publishes fluctuation of 4.0 ± K, larger than its own uniformity of 3.5 ± K, while UF V 500 runs the other way, 1.5 against 2.5. The two parameters do not move together and neither predicts the other.
测点范围不同,不只是数字大小不同。中国标准取全测点最大值,最坏的那个点决定结果;品牌数据表常报单个控制探头点。同一台机器,两种算法可以差出一倍以上,而两个数字看起来是同一个指标。另有一个坑:部分文档把 temperature stability 当作 fluctuation 的同义词,另一些里它指长期漂移(年漂移),两者不可互换。 还有一个反直觉的实例:Binder UF V 350 的波动度(4.0 ± K)**大于**它自己的均匀度(3.5 ± K),而 UF V 500 正好相反(1.5 对 2.5)。两个参数不是同向变化的,不能拿一个去推另一个。
Ask before comparing看数字前先问
先问:是全测点最大还是单点?采样时长多久?有没有排除化霜?只给一个 ±值的,按「单点、未说明」处理,不得直接与全测点数据比较。
Sources 来源- NIFDC (National Institutes for Food and Drug Control) · 医用冷链设备术语(公开征求意见稿)3.37 · 原文Tier 4 · standard or regulator
- Binder · Model UF V 500 technical data(官方产品页) · 原文Tier 2 · manufacturer technical document
- PHCbi · Ultra-Low Temperature Freezers: By the Numbers · 原文Tier 2 · manufacturer technical document
- Thermo Fisher Scientific · 超低温冰箱的特点(品牌 ULT 特性页,中文版) · 原文Tier 2 · manufacturer technical document
- Thermo Fisher Scientific · TSE240V Technical Specification Sheet(品牌官方数据表) · 原文Tier 3 · industry guidance
Setpoint
Aligned设定温度
A -86 °C freezer means the lowest setpoint the machine can hold, not a chamber that is -86 °C at every point.
What it measures定义什么
The target value the controller works to hold. It is a control target, not the actual temperature at any position in the chamber.
控制器要去维持的那个目标值。它是一个控制目标,不是箱内任意一点的实际温度。
How it is measured怎么测
Set on the controller. Actual temperature moves about the setpoint: the movement over time is described by fluctuation, the spread over space by uniformity.
由控制器设定。箱内实际温度会在设定点上下波动,波动幅度由 fluctuation 描述,空间偏离由 uniformity 描述。
Where the numbers part口径差异
Definitions agree. The risk sits in the wording: a -86 °C freezer in a brochure names the lowest achievable setpoint, not the temperature at every position. Actual temperature lies in a band around it, described by uniformity and fluctuation.
各家定义一致。真正的风险不在定义,而在表述:宣传里的「-86 °C 冰箱」指的是最低可达设定点,不是箱内每处都 -86 °C。
Sources 来源- Thermo Fisher Scientific · Smart Note: Understanding Pull Down · 原文Tier 2 · manufacturer technical document
- Thermo Fisher Scientific · TSE240V Technical Specification Sheet(品牌官方数据表) · 原文Tier 3 · industry guidance
Pull-down time
Differs降温时间
Marketing sells it as a headline; the manufacturer's own technical note says it describes a rare event and can even represent wasted energy.
营销材料拿它当卖点,Thermo Fisher 自己的技术文档说它不代表日常性能,甚至可视为浪费能耗。
What it measures定义什么
Time for an empty chamber to fall from ambient to setpoint. It describes a one-off event at first installation or after a manual defrost, not day-to-day performance.
空箱从环境温度降到设定温度所需的时间。它描述的是首次安装或手动除霜之后的一次性事件,不是日常使用中的性能。
How it is measured怎么测
Start from ambient with an empty chamber and record the time to reach setpoint. The starting ambient temperature has to be stated alongside the figure.
空箱、常温起动,记录回到设定点的时间。测试时的环境温度必须一并给出。
Where the numbers part口径差异
The disagreement is not about the number but about whether it indicates quality at all. Marketing treats a fast pull-down as a selling point, while a Thermo Fisher technical note states that initial pull-down is a rare event, at first installation or after a manual defrost, with no samples inside, and that a fast initial pull-down does not necessarily indicate superior performance and can be regarded as wasted energy. Two documents from one company, one promoting and one demoting the same figure: which is why the source has to be named.
这里的分歧不在数值,在「该不该用它判断好坏」。营销材料普遍把降温快当卖点;而 Thermo Fisher 的技术文档明确说:初次降温是罕见事件(首次安装、手动除霜后),此时样品并不在箱内,「快速初次降温不一定说明性能优越,反而可以视为在浪费能耗」。同一家公司的两份材料,一份主打、一份降级,这就是必须列明出处的理由。
Ask before comparing看数字前先问
把它当安装参数看,不要当日常性能看。要判断日常性能,看 door opening recovery。
Sources 来源- Thermo Fisher Scientific · Smart Note: Understanding Pull Down (SNULTDOR 1119) · 原文Tier 2 · manufacturer technical document
- Binder · Model UF V 500 technical data(官方产品页) · 原文Tier 2 · manufacturer technical document
- Thermo Fisher Scientific · 超低温冰箱的特点(品牌 ULT 特性页,中文版) · 原文Tier 2 · manufacturer technical document
Door opening recovery
Differs开门恢复时间
Fifteen seconds of door-open and sixty seconds produce two different figures, and brochures rarely state which was used.
开门 15 秒和开门 60 秒测出来的恢复时间,是两个数字。宣传页很少写开门时长。
What it measures定义什么
Time for chamber temperature to return to setpoint after the door has been open. It sits closer to daily use than pull-down: an ultra-low freezer is typically opened two to four times a day, twenty in a busy laboratory.
门开过之后,箱内温度回到设定点所需的时间。它比降温时间更接近日常:一台超低温冰箱通常每天开门 2–4 次,繁忙实验室可达 20 次。
How it is measured怎么测
Record how long the door stayed open, how far temperature moved during that time, and how long it took to return to setpoint or into a stated tolerance band. All three have to be given, otherwise the figure cannot be compared.
记录开门时长、开门期间的箱温变化、以及温度回到设定点(或某个容差带内)所需时间。三项必须同时给出,否则数字不可比。
Where the numbers part口径差异
Different door-open durations are simply not comparable. The ENERGY STAR sequence opens the door for 15 seconds, six times an hour; brochures often say recovery after 60 seconds. On the same machine, a 60-second recovery runs two to three times the 15-second figure. The recovery criterion is often left unstated too: return to setpoint or return into a tolerance band can differ by half the time.
开门时长不同,数字根本不可比。ENERGY STAR 的测试序列是「每小时开门 15 秒、连续 6 次」;品牌宣传里常见「开门 60 秒后恢复」。同样的机器,60 秒开门的恢复时间是 15 秒的两三倍。另外「恢复」的判据各家常不写清:是回到设定点,还是回到某个容差带内,两者能差出一半时间。
Ask before comparing看数字前先问
必须问清:开门多少秒?恢复到什么判据?空箱吗?环境温度多少?四项缺一项,这个数字只能当参考,不能进对比表。
Sources 来源- Thermo Fisher Scientific · Smart Note: Understanding Pull Down · 原文Tier 2 · manufacturer technical document
- ENERGY STAR · ULT-ECO Freezers: By the Numbers · 原文Tier 4 · standard or regulator
- Technology Networks 访谈(Thermo Fisher 的 Andrew Dunn) · A Revolution in Cold Storage · 原文Tier 5 · secondary source
- Thermo Fisher Scientific · 超低温冰箱的特点(品牌 ULT 特性页,中文版) · 原文Tier 2 · manufacturer technical document
- Thermo Fisher Scientific · TSE240V Technical Specification Sheet(品牌官方数据表) · 原文Tier 3 · industry guidance
Gross volume / Net (effective) volume
Differs总容积 / 有效容积
The litre figure on a brochure rarely says gross, net or nominal. The three differ by more than ten per cent, and what decides sample count is box capacity.
彩页上的 L 数常常不说自己是毛容积还是有效容积。两者能差十几个百分点,而真正决定能放多少样本的是盒数。
What it measures定义什么
Gross volume is all the space enclosed by the inner walls with nothing subtracted. Net, or effective, volume is the space that can actually be loaded, after evaporator, air ducts, shelves and the height above the stacking mark are deducted.
总容积是内壁围起来的全部空间,不扣任何内部件;有效容积是真正能放东西的空间,要扣掉蒸发器、风道、搁架、以及堆叠标记以上的高度。
How it is measured怎么测
Gross volume from the internal geometry. Net volume from the width and depth of the horizontal shelf and the height limited by functional channels, components or the stacking mark, per DIN 13221.
总容积按内壁几何尺寸算。有效容积按水平搁板的宽与深、以及由功能通道、组件或堆叠标记限定的高度算(DIN 13221)。
Where the numbers part口径差异
The definitions agree; the labelling practice does not. A brochure often states neither gross nor net, and a third term, nominal, is also in use, with more than ten per cent between them. What decides how many samples fit is not any of the three but box capacity.
标准里的定义其实是一致的,分歧出在标注实践:厂家的彩页容积常常既不说明是 gross 也不说明是 net,第三种叫法是 nominal(标称),三者可以差出百分之十几。真正决定能放多少样本的不是任何一个容积,而是盒子数量(box capacity)。
Ask before comparing看数字前先问
看到 L 数先问:gross、net 还是 nominal?再看同一页有没有给 2 英寸盒数量。没给盒数的容积,不能用来估算样本量。
Sources 来源- DIN (German Institute for Standardization) · DIN 13221:2016-08 Laboratory refrigerators and freezers — Terminology, requirements, testing · 原文Tier 1 · manufacturer standard
- IEC · IEC 62552(家用制冷器具,容量测量方法被实验室冰箱行业沿用)Tier 1 · manufacturer standard
Cascade refrigeration
Aligned复叠式制冷
Two circuits in series: the high stage cools the condenser of the low stage, which is what lets the low stage go below -40 °C.
What it measures定义什么
Two refrigerant circuits in series: the high stage cools the condenser of the low stage, which is what allows the low stage to go below -40 °C. This is the usual architecture at the -86 °C level.
两个制冷回路串联:高温级把低温级的冷凝器冷下来,低温级才能降到 -40 °C 以下。这是 -86 °C 级别的主流构型。
How it is measured怎么测
Not applicable: a system architecture, not a measured value.
不适用(系统构型,非测量值)
Where the numbers part口径差异
Descriptions agree. What has to be separated is cascade from single-stage compression with other aids: single-stage normally stops near -40 °C, so a claim of -86 °C without cascade calls for a question about the circuit.
各家描述一致。要区分的是「复叠」与「单级压缩 + 其他辅助手段」:单级压缩通常止步于 -40 °C 左右,声称 -86 °C 而不用复叠的,要问清构型。
Sources 来源- PHCbi · ULT freezer brochure — Cascade Cooling System · 原文Tier 2 · manufacturer technical document
Effective data collection sample
Differs有效的数据采集样本
The Chinese standard's definition of which readings count. Sampling by this rule and sampling by brand habit do not produce comparable figures.
这是中国标准对「什么数据才算数」的界定。按它取样与按品牌惯例取样,结果不能直接比。
What it measures定义什么
The Chinese standard's definition of which readings count: once the environment is stable and the unit has reached steady state, a whole-number multiple of control cycles closest to three hours of continuous running, excluding defrost, with at least one reading per point per minute. Only these readings may be used to compute fluctuation and uniformity.
中国标准里对「什么数据才算数」的界定:环境稳定、设备进入稳定状态后,连续运行最接近 3 小时的正整数倍周期(不含化霜),且每个测点每分钟至少取一次数。这些数据才可用于计算波动度与均匀度。
How it is measured怎么测
Not applicable: a sampling rule, not a measuring method.
不适用(这是取样规则,不是测量方法)
Where the numbers part口径差异
International brand documents usually say only that measurement was taken after stabilisation, without so specific a rule. Figures taken to the Chinese standard are therefore not necessarily comparable with brand figures. Neither is stricter; the sampling conditions differ.
国际品牌文档通常只说「测试稳定后进行」,不给这么具体的取样规则。所以按中国标准测出来的数,与国际品牌的数不一定能直接对比——不是谁更严,是取样条件不同。
Ask before comparing看数字前先问
若一方按此规则取样、另一方未说明,两者不可直接比较;需向未说明方索取测试报告。
Sources 来源- NIFDC (National Institutes for Food and Drug Control) · 医用冷链设备术语(公开征求意见稿)3.36 · 原文Tier 4 · standard or regulator
Steady state
Aligned稳定状态
The threshold for starting to measure: no point moves by more than 0.5 K over two hours.
What it measures定义什么
The threshold for saying measurement may begin: readings at every point, taken at the same moment in the control cycle, change by no more than 0.5 K over two hours.
判断「可以开始测了」的门槛:箱内各测点在温度控制周期同一时刻的读数,在 2 小时内变化不超过 0.5 K。
How it is measured怎么测
Repeat the measurement at the same point in the control cycle and compare the change over two hours.
在温度控制周期的相同时刻重复测量,比较 2 小时内的变化量。
Where the numbers part口径差异
This is the quantified threshold the Chinese standard gives. Brand documents usually give no figure and say only after stabilisation. Before comparing two sets of data, confirm that both mean the same thing by stable.
这是中国标准给出的量化门槛。品牌文档通常不给具体阈值,只说「待稳定后」。所以对比两家的数据时,要先确认双方对「稳定」的定义是否一致。
Sources 来源- NIFDC (National Institutes for Food and Drug Control) · 医用冷链设备术语(公开征求意见稿)3.40 · 原文Tier 4 · standard or regulator
ENERGY STAR (ULT freezers)
Aligned能源之星(超低温冰箱)
Its value is not the energy label but independent testing under conditions identical for every participant, which makes it one of the few sources comparable across makes.
What it measures定义什么
A voluntary efficiency programme run by the US EPA and DOE. Its value is not the energy-saving label but the requirement of independent third-party testing under conditions identical for every participating manufacturer, which makes it one of the few sources that can be compared across makes.
美国 EPA 与 DOE 设立的自愿性能效认证。价值不在「省电」这个标签,而在它强制第三方独立测试,且测试条件对所有参与厂家一致,所以是少数可以横向比的数据来源。
How it is measured怎么测
Tested independently by a certification body, with manufacturer self-test data not accepted. Total energy over 24 hours at both the -70 °C and -80 °C setpoints, 24 °C ambient, starting from steady state, door opened for 15 seconds every hour for six hours.
由认证机构独立测试,厂家自测数据不被接受。24 小时总能耗,取 -70 °C 与 -80 °C 两个设定点;环境 24 °C;稳定后开始,每小时开门 15 秒、连续 6 小时。
Where the numbers part口径差异
The certification itself is consistent. What to watch is that in-house testing commonly uses 20 °C or 30 °C ambient rather than the 24 °C of ENERGY STAR, and both consumption and recovery deteriorate markedly at 30 °C, so in-house and certified figures must not be mixed.
认证本身各家一致。要注意的是:品牌内部测试常用 20 °C 或 30 °C 环境,与 ENERGY STAR 的 24 °C 不同;而 30 °C 环境下的能耗与恢复数据会明显变差,所以品牌自测与认证数据不能混着比。
Ask before comparing看数字前先问
确认数字来自 ENERGY STAR 认证报告还是厂家自测;两者不可混比。
Sources 来源- ENERGY STAR · ULT-ECO Freezers: By the Numbers · 原文Tier 4 · standard or regulator
2-inch box / box capacity
Differs2 英寸冻存盒 / 盒容量
480 boxes by 81 places is 38,880 vials; 480 boxes by 100 places is 48,000. One sentence, holds 480 boxes, hides nearly ten thousand vials.
480 盒 ×81 格 = 38,880 管;同样 480 盒 ×100 格 = 48,000 管。同一句「可存 480 盒」,实际管数差近一万。
What it measures定义什么
A standard cryo box is 50 mm high, two inches, usually with 81 places, 9 by 9, for vials up to 13 mm, or 100 places, 10 by 10. The box count a manufacturer publishes is the number of storage units, not of sample vials.
标准冻存盒高 50 mm(2 英寸),常见规格为 81 格(9×9,适配管径 ≤13 mm)或 100 格(10×10)。厂家标的「可放多少盒」是收纳单元数,不是样本管数。
How it is measured怎么测
Counted by box positions on shelves or in drawers, independent of the grid inside the box and of vial diameter.
按层架或抽屉的盒位计数,与盒内格数、管径无关。
Where the numbers part口径差异
Box count and vial count are often used interchangeably. 480 boxes by 81 places is 38,880 vials; 480 boxes by 100 places is 48,000. The same sentence, holds 480 boxes, hides nearly ten thousand vials. When only a box count is given, the sample count has to be multiplied out and the grid assumption stated.
盒数与管数常被混着说。480 盒 ×81 格 = 38,880 管;同样 480 盒 ×100 格 = 48,000 管。同一句「可存 480 盒」,实际管数差近一万。厂家给的是盒数时,样本量必须自己乘,并注明格数假设。
Ask before comparing看数字前先问
把盒数换算成管数时必须写清格数假设(81 或 100)与管径,否则数字无意义。
Sources 来源- PHCbi · ULT freezer brochure · 原文Tier 2 · manufacturer technical document