Specifying a 2 to 8 °C pharmaceutical refrigerator

The storage band is written into the buyer's licence or the buyer's SOP, so it is not a choice. Everything that decides whether the cabinet passes an inspection is chosen afterwards: how the air moves, where the temperature is measured, what happens when the door opens, and whether the record survives a hard disk.

A pharmacy refrigerator is judged on one thing: whether the products inside stayed within the band the whole time, and whether the buyer can prove it. Performance under a full load, alarm behaviour during a power cut, and the temperature record are the three questions an inspector will ask. They are also the three that are easiest to leave undefined at the specification stage.

Usable volume, not the number in the model name

The figure in a model name is a gross figure. Shelves, drawer fronts, the door liner and the evaporator housing all take space inside it, and the gap is not small.

Ask for the usable volume and the shelf count, then count the buyer's owns stock against them. Do that with a third of the space left over, for two reasons: an overfilled cabinet has no air path, and an empty cabinet has no thermal mass, so it recovers slowly and unevenly after the door closes.

The 2 to 8 °C range in HYSI's catalogue runs from 150 L through 395, 455, 635 and 795 L to 1150 L. Two further models are listed without a volume, and their dimensions, shelving and cryovial capacity are confirmed per model and sent by email.

Where the temperature is measured

Cold air is denser than warm air, so a single-probe cabinet always has a gradient. The number on the display tells the buyer what the probe sees, not what the top shelf sees.

Two questions follow. Where is the control probe mounted, and where does the monitoring probe go. If the monitoring probe is sited next to the control probe, the two will agree and tell the buyer very little about the load.

Before the buyer write an acceptance test, map the cabinet with a full load: band and corners, door closed, at the point in the day when the door is opened most. That map is the document the buyer's inspector will asks for.

Air distribution and recovery

A cabinet that holds temperature well may still recover badly. Recovery is what matters in a busy dispensary, where the door opens twenty times an hour.

Ask how long the cabinet takes to return to band after a door opening of a stated duration under a stated load. It is a measure the buyer can accept or reject on, and it is the one that explains most complaints after installation.

Shelf layout is the lever the buyer control. Leave the air path clear, and do not stack stock in front of the return.

Alarms, records and the audit

Three alarms matter: temperature out of band, door left open, and mains failure. Each one needs a defined action, not just a buzzer.

The record matters more than the alarm. A data logger that exports to a file the buyer can reads years later is worth more than a display that shows a minimum and maximum. Ask how the log is exported, how often it samples, and how long it retains.

Where a product requires a validated chain, the log is part of the release paperwork. Decide where it lives before the buyer decides which cabinet to buys.

Door format, room and power

A glass door lets staff read a label without opening the cabinet, and it puts more heat into the room. A solid door insulates better. Which one wins depends on how often the cabinet is opened and how much cooling the room already carries.

Then measure the route. A 1150 L cabinet has to pass a doorway, turn a corridor corner and stand on a floor that can take it. HYSI checks the door and corridor clearance with the buyer at the same time as the dimensions, because the answer changes the model.

Finally, power. HYSI's cold-chain range is built to a voltage and frequency confirmed in writing before orders, with the plug configuration to matches the buyer's market.

What to confirm before the buyer orders

Item What to pin down Why it decides the purchase
Usable volume and shelving Internal volume, shelf count, drawer configuration, cryovial capacity Gross volume is what the name says; usable volume is what holds the buyer's stock
Temperature measurement Probe position, monitoring point, mapping under full load An inspector reads the record, not the display
Recovery Return to band after a door opening, at a stated load Explains most day-to-day complaints
Alarms and logging Out-of-band, door open, mains failure; export format and retention Turn a buzzer into a document the buyer can hand over
Dimensions and access Exterior dimensions, door swing, corridor and doorway clearance A cabinet that cannot be moved in is not a delivered cabinet
Power Voltage, frequency, rated power, plug Confirmed in writing before order, to the buyer's market standard

What to send

Send the band the buyer has to holds, the products and their packaging, how many units of each the buyer stores, the door-opening traffic, the room dimensions and the power available.

The manufacturer answers with the model, the part number, the shelving layout and the dimensions, with documentation listed as its own item. Where the buyer's figures leave two models in play, the page names the measurement that separates them.

Related models

Answers before the buyer specifies

Why does the display disagree with my data logger?

Because they measure in different places. A probe mounted near the evaporator reads colder than the top shelf. Ask where the control probe sits, and map the cabinet under a full load before the buyer write the acceptance test.

Do I need a glass door?

A glass door lets a pharmacist see stock without opening the cabinet, and it rejects more heat into the room. Whether that is a gain depends on how often the door is opened and how much cooling the room already has.

What happens during a power failure?

A loaded cabinet holds its temperature longer than an empty one, which is one reason to plan shelf space generously. Ask for the warm-up time under a stated load, and treat the alarm and the backup supply as separate decisions.

Can one cabinet hold vaccines and reagents together?

Only if the two products share a storage band and a segregation rule. Where they do not, the usual answer is two cabinets rather than one larger cabinet.

Which volumes can the buyer supplies?

The 2 to 8 °C range in HYSI's catalogue runs from 150 L through 395, 455, 635 and 795 L to 1150 L. Two further models are listed without a volume, and their figures are confirmed per model by email.