Both machines hold a rack of vials and remove solvent from all of them at the same time. One sweeps the surface with dry gas, the other lowers the pressure above it. The right answer follows from the solvent, the vial and the analyte, in that order.
The difference between the two routes is where the energy for evaporation comes from. Blowdown replaces the saturated vapour above the liquid with dry gas, so the liquid keeps evaporating at whatever temperature the block holds. Parallel evaporation lowers the pressure instead, so the liquid boils below its normal boiling point and the bath only has to make up the heat of vaporisation.
Start here, because it is the least negotiable of the three.
If the buyer's samples arrive in 2 mL autosampler vials, a parallel evaporation concentrator runs 100 of them at once from a water bath at room temperature to 80 °C, with orbital shaking of 0 to 600 rpm and a vacuum down to 10 mbar. The rack is the instrument.
If the buyer's samples are in 10 to 50 mL tubes, a nitrogen blowdown concentrator takes 24 of them, with gas flow adjustable from 0 to 12 L/min and a block holding room temperature plus 5 °C up to 100 °C.
Converting between the two formats means buying new racks, and the racks cost more than most laboratories expect. Check the vial the buyer already buys before anything else.
Dichloromethane, methyl tert-butyl ether and hexane leave a vial quickly under gas flow alone. The drying step is short and the block temperature can stay low, which suits analytes that would prefer not to be warmed at all.
Water, dimethyl sulfoxide and dimethylformamide are the other end. Warming alone takes a long time, and the vacuum does more useful work, because pressure rather than temperature is the variable that moves those boiling points.
A single rule follows from this: the higher the boiling point, the more the vacuum earns its place.
Both routes let the buyer holds a temperature below 40 °C, which is where most thermolabile works lives. The difference is the gas.
A nitrogen blanket displaces oxygen from the headspace, so a blowdown concentrator also protects analytes that oxidise. A vacuum chamber can do the same job with a backfill cycle, but that has to be asked for rather than assumed.
Bumping is the failure mode to design around on the vacuum route. Liquid thrown into the gas path is sample lost, not solvent lost. Throttling the vacuum, lowering the bath and keeping the shaker running all reduce it.
Blowdown vents the solvent to the fume hood. That is acceptable at low volume and awkward when the buyer is running hundreds of millilitres a day.
Parallel evaporation can recover the solvent when a cold trap is fitted. On the model in HYSI's catalogue the trap is an option and is quoted as a separate line, so the buyer can sees what the recovery costs against what the solvent costs the buyer.
| Question | Nitrogen blowdown | Parallel evaporation |
|---|---|---|
| Driving force | Dry gas sweep over the surface | Reduced pressure above the liquid |
| Vial format in HYSI's catalogue | 24 × 10 to 50 mL tubes | 100 × 2 mL vials |
| Temperature control | Room temperature +5 to 100 °C | Bath from room temperature to 80 °C |
| Agitation | Gas flow only | Orbital shaking, 0 to 600 rpm |
| Solvent recovery | Vented to the hood | Cold trap available as an option |
| Suits | Volatile solvents, oxygen-sensitive analytes, larger tubes | Large batches of small vials, water and high-boiling solvents |
Send the vial or tube the buyer already buys, the solvent, the starting volume, the temperature ceiling from the buyer's method, and the number of samples per batch. That is enough to choose the route and the model.
The manufacturer answers with the model, the part number, the rack or tray that matches the buyer's vessel, and the delivery term. Where a cold trap or an extra rack is needed, it is listed as its own item.
It can, but many methods stop just short of dryness, because an analyte that has gone completely dry can adsorb on the vessel wall. Stop at the residual volume the buyer's method defines.
A bump throws liquid into the gas path, so the loss is the buyer's sample and not the solvent. It is controlled by throttling the vacuum, lowering the bath temperature, or shaking the vials so boiling stays gentle.
For volatile solvents the two are close, and the gas flow is easier to tune. For water and high-boiling solvents the vacuum route pulls ahead, because reducing pressure does more than warming the surface.
The buyer needs one if the solvent has to be recovered, or if the pump would otherwise see the vapour. On the model in HYSI's catalogue the trap is an option, quoted separately so the buyer can sees the cost.
Nitrogen is the usual choice because it is inert and dry. Flow is set per position rather than per instrument, because a vial with less liquid needs less gas to stay below the surface.
Questions about what a figure means, where it came from, or why a model is listed as it is are answered by the index. Questions about price, delivery or configuration are answered by the manufacturer; the model page carries the number to call.