Choosing a microscopy contrast method

Most mis-bought microscopes are bought on magnification. A specimen that transmits light without absorbing it looks empty at any magnification, and no amount of extra optics fixes that. Contrast is what shows the buyer the sample, so contrast is where the specification starts.

A contrast method turns one property of the specimen into visible difference. Absorption, scattering, refractive index, birefringence and fluorescence are five different properties, and the technique that reads one of them will show nothing of the others. Choose the property that the buyer's specimen actually has.

Contrast comes from the sample

Start by describing what makes the buyer's specimen different from the medium around it. A stained tissue section absorbs light, and the difference is already there. A living cell in buffer is almost transparent, so the difference has to be created by the optics.

This is the question to answer before a model is named. When a laboratory asks for a higher magnification because the image looks faint, the tool being asked for is usually the wrong one.

Brightfield and darkfield

Brightfield is the default and the one every stand does well. Light passes through the specimen and the absorbing parts appear darker against a bright background. Stained sections, smears and any preparation with natural colour belong here.

Darkfield blocks the direct light so that only scattered light reaches the objective. The background goes black and edges, outlines, particles and fine transparent structures stand out against it. It suits unstained suspensions, small particles and the search for something faint in a clean field.

One metallurgical stand in HYSI's catalogue carries both, which is useful when the same bench has to inspect both prepared and unprepared samples.

Phase contrast, and the DIC slot

Phase contrast converts differences in refractive index into differences in brightness, which is what makes unstained living cells visible. It needs an annular stop in the condenser and objectives matched to it, so the objective set has to be specified with the method rather than after it.

Differential interference contrast gives a similar result with a pseudo three-dimensional appearance and a thin optical section, which can be preferable for thick or highly detailed specimens. The optical components are more expensive, which is why a stand may be bought with a slot and the prism added later when the application justifies it.

An inverted stand for cell culture is the usual home for this work, since the vessel has to sit above the objective.

Polarising light and fluorescence

A polarising microscope places the specimen between two polarising filters and shows birefringent material against a dark background. Crystals, minerals, fibres, starch and a range of polymers respond to it, and a rotating stage with a graduated scale lets the buyer measures extinction angles rather than just look.

Strain-free objectives matter here, because an objective that introduces its own birefringence adds its own signal to the measurement. HYSI's polarising range is specified with strain-free plan objectives for that reason.

Fluorescence is the opposite approach: the specimen is labelled, and the microscope filters out the excitation light while collecting the emission. It is the most sensitive of the methods and the most dependent on the filter set, so the filter turret and its band selection have to match the fluorophores in the buyer's protocol.

The stand decides the working distance

Once the contrast method is settled, the stand has to match the physical form of the sample.

An upright stand views a slide from below, with the specimen on top of the stage. It is right for sections, smears and most prepared samples, and it is the cheaper way to reach high magnification.

An inverted stand puts the objective under the stage so the vessel sits above it. This is the stand for flask and plate culture, and the working distance has to clear the thickness of the plastic. It is also the stand used for micromanipulation and for observing sediments.

A stereo stand gives two angled light paths, so the image has depth and the working distance can be tens of millimetres, which leaves room for tools. Zoom ratios in HYSI's catalogue runs from 6:1 to 6.7:1, with working distances from 110 mm to 301 mm depending on the objective fitted.

A metallurgical stand works in reflected light because the specimen is opaque. Polished mounts, coatings and machined surfaces belong here, and the stage size grows with the part being inspected. The large-stage model carries a 420 mm by 400 mm stage with 200 mm by 200 mm travel.

Industrial measuring and probe station stands are the last group: long working distances, apochromatic objectives for geometric accuracy, and stages built to hold a wafer or a fixture rather than a slide.

Magnification, field of view and the camera port

Total magnification is the objective times the eyepiece, which means the number on a specification sheet is easy to inflate and tells the buyer little on its owns.

The field number is the more useful figure. It describes how wide a field the eyepiece shows, and HYSI's range runs from 16 to 25 across the eyepieces HYSI supplies. Two stands that both claim 1000X feel completely different to use, because one shows a field of 25 mm and the other a field of 16 mm.

If the work has to be documented, choose a trinocular head and check the beam split. HYSI's research stands offer selectable ratios of 100:0, 20:80 and 0:100, so light can go to the eye, to the camera, or to both in a fixed proportion.

Where measurement matters more than visual work, a digital stand takes the place of the eyepieces. The ultra-depth model in HYSI's catalogue covers 30x to 10000x with 0.1 µm measurement resolution, which is outside the range of a conventional optical stand.

Method against sample

Method Property it reads Suits What it needs
Brightfield Absorption Stained sections, smears, coloured preparations Nothing beyond the stand
Darkfield Scattering Unstained suspensions, particles, outlines A darkfield stop in the light path
Phase contrast Refractive index Living cells and thin transparent specimens Annular stop plus matched objectives
Polarising Birefringence Crystals, minerals, fibres, polymers Two polarisers, a rotating stage and strain-free objectives
Fluorescence Emission from a label Labelled cells, tissue and particles Excitation source and a filter set matched to the label

What to send

Send the specimen and how it is prepared, the property the buyer is trying to sees, whether the sample is living, and the vessel if it is not a slide. Add the measurements the buyer has to take and whether they have to be recorded.

The manufacturer answers with the stand, the contrast method, the objective sets and the part number. Where two methods would both work, the page states which costs less to buy and which is easier to extend later.

Related models

Answers before the buyer specifies

What is the difference between brightfield and darkfield?

Brightfield lights the sample from below and shows what absorbs light. Darkfield blocks the direct beam and shows what scatters it, so edges, particles and outlines stand out against a dark background while the interior stays empty.

Do I need phase contrast or a DIC slot?

Phase contrast is the common choice for unstained living cells, and it needs a matching objective set. A DIC slot in the stand lets the prism be added later where the application calls for it, without replacing the microscope.

Why does the field number on the eyepiece matter?

It sets how much of the specimen the buyer sees at a given magnification. A 10X eyepiece with a field number of 23 shows a wider field than one at 18, which is why two microscopes with the same magnification can feel completely different to work with.

Can I add a camera later?

Choose a trinocular head now if the buyer intends to. The beam split ratios on HYSI's research stands are selectable, so light goes either to the eye, to the camera, or to both in a fixed proportion.

When is an inverted microscope the right stand?

When the sample is in a flask, a dish or a well plate and the buyer observe it from below. The objective looks up through the base of the vessel, so the working distance has to clear the plastic.