A concentrator holds the buyer's sample against the wall of a spinning chamber while vacuum and mild heat pull the solvent off. Get the family wrong and the buyer either lose the analyte or spend the afternoon waiting. Below is the order in which HYSI settles a specification, and the numbers HYSI asks the buyer for.
A centrifugal vacuum concentrator is not a centrifuge and not a rotary evaporator. It spins the liquid to the bottom of the vessel so the surface keeps renewing while a pump lowers the pressure above it. The specification follows from the solvent, the vessel and the temperature the buyer's sample tolerates.
Water and dilute buffers are the easy case. A chamber that reaches ambient to +70 °C is usually enough, and no trap is needed.
Dichloromethane, acetonitrile and methanol are the next case. The vapour has to condense somewhere before it reaches the pump, so a cold trap is part of the instrument, not an accessory. A trap at −45 °C condenses most of it.
Acid digests are the third case, and they are the one that ends projects. The vapour is acidic. Chamber, rotor, seal, tubing and pump gas path all have to tolerate it. A standard diaphragm pump with an aluminium head will not see the end of an acid programme.
This is the question HYSI asks first. It removes more models than any other.
The buyer is not pelleting cells. The buyer needs enough force to keep the liquid from climbing out of the tube, and no more.
Models in the HYSI catalogue runs 800 to 2,000 rpm, or 1,500 rpm on the automated frames, with a maximum RCF of 530 × g. That figure looks low next to a preparative centrifuge. It is deliberate: above roughly 500 × g the residue starts to compact, and a compacted residue is slow to redissolve.
So ask how the buyer will redissolve. Pipetting up and down favours low RCF. Vortexing tolerates more.
Ambient to +70 °C. Aqueous and volatile solvents where heat is not a risk.
Ambient +5 °C to +100 °C. A floor above room temperature, and a ceiling high enough for water.
−10 to +100 °C. Heat-labile work: proteins, enzymes, nucleic acid extracts. The refrigerated frame holds the setpoint to ±1 °C.
Evaporation cools the sample. A chamber reading +30 °C on the display can be colder at the tube. If the buyer's protocol says do not exceed 30 °C, asks what the instrument holds under full vacuum rather than at rest, and put the answer in the orders as an acceptance item.
Two numbers matter. The control range sets how finely the buyer can throttle the vacuum. The ultimate vacuum sets how dry the buyer can get.
Catalogue control ranges run 0.1 to 99.9 mbar, or to 1,000 mbar on the solvent workstations that also have to handle high-boiling solvents. Ultimate vacuum sits at or below 0.2 to 0.3 mbar.
Then the vapour has to go somewhere. Run a chlorinated solvent without a trap and the pump is the part that fails first, usually within a year.
Throughput comes from the rotor, not the chamber. Formats covered include 40 × 2 × 1.5 mL, 40 × 2 × 2 mL and 24 × 2 × 5 mL, plus plate rotors on the automated frames.
Two habits save trouble later. Leave headspace, because a 1.5 mL tube filled to the rim will bump. And match the rotor to the consumable the buyer already buys, rather than moving the laboratory onto new consumables.
Write down each wetted part between the sample and the exhaust: chamber, rotor, seal, tubing, trap, pump head. Then check each one against the solvent. For acid work, specify a chamber that resists it and a pump with a PTFE-lined gas path.
| Family | Speed / RCF | Temperature | Vacuum control | Typical work |
|---|---|---|---|---|
| Acid-resistant, refrigerated | 800–2000 rpm, 530 × g | −10 to +100 °C, ±1 °C | 0.1–99.9 mbar | Acid digests, trace metal preparation |
| Acid-resistant, ambient-plus | 800–2000 rpm, 530 × g | Ambient +5 to +100 °C, ±1 °C | 0.1–100 mbar | Inorganic samples without a heat-labile step |
| Solvent evaporation workstation | 800–2000 rpm, 530 × g | Ambient to +80 °C | 0.1–1000 mbar | High-boiling solvents, cold trap at −45 °C |
| Automated frame | 1500 rpm | Ambient to +70 °C, ±2 °C | 0.1–99.9 mbar | Plate formats, integration into an automation line |
Send the solvent or matrix, the starting volume and the vessel, the temperature ceiling the buyer's protocol sets, and the criterion the buyer has to meet. That is enough for HYSI to answer with a specification rather than a brochure.
The manufacturer answers with the model, the part number and the delivery term. Where nothing in the catalogue meets the criterion, the page says so rather than leaving the buyer to find out after ordering.
Usually not. Chlorinated solvents need a cold trap and a corrosion-resistant gas path. Have the trap specified with the instrument rather than added afterwards.
Deep enough to remove the solvent at a temperature the buyer's sample tolerates. Water needs a lower vapour pressure than methanol. The control range matters as much as the ultimate figure, because the buyer will throttle it.
It should. Rotors are inexpensive to change, but moving a laboratory onto new consumables costs more over time. Send the vessel dimensions and the rotor is matched to them.
Ask what the chamber holds under full vacuum, not at rest. Evaporative cooling takes the sample below the setpoint. If the buyer's method has a ceiling, make it a written acceptance item.
They are quoted as separate line items, so the buyer can sees what each one costs. The pump gas path is specified to the solvent the buyer uses, not to the catalogue.